- 67 -
Fig. 54. Statistics display content including a series of measurements
1. Mass currently placed on the weighing pan
2. Number of measurement in a measuring series
3. Sum of all completed measurements in a series
4. Average mass of all completed measurements in a series
5. Minimum saved mass in a carried out measuring series
6. Maximum saved mass in a carried out measuring series
7. The difference between the minimum and maximum saved mass in a measuring
series
8. Precisely calculated standard deviation value
9. Variance factor value
10. Measuring unit [g]
11. Working mode name
Statistics mode features a special option enabling quick calculating of statistics from a
measuring series with a possibility of generating reports from the completed calculations.
The user can declare the content of a header and footer in a printout of statistical report.
The header <PRINTOUT 1>, and the footer <PRINTOUT 2> are settable in non-standard
printouts.
In order to correctly use the option, the user has to create templates 1 and 2 using the non-
standard printouts (see point 19.2). As the templates are ready, the option can be enabled
in mode settings.
Means of operation:
Press key to print the header (PRINTOUT 1)
Carry out a measuring series (set all options of the GLP menu to NO), and accept
each measurement result by pressing <ENTER> key
On completing a measuring series press F key to print a statistics from the series
In order to continue the measuring series accept it by pressing <ENTER> key or
Press key to print a footer (PRINTOUT 2)

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43