- 28 -
Apart from information set in menu group the report contains: Mass of adjustment weight
stored in balance’s memory from last carried out adjustment (description: Old:) Mass of
adjustment weight determined in current adjustment process (description: Adjustment:)
Adjustment deviation, i.e. difference between the two mass records (description:
Deviation:).
Fig 17. An example of a report from adjustment process
8. DETERMINING CONTENT OF A PRINTOUT FOR GLP PROCEDURES
Menu P2 GLP is group of the parameters which enables declaring variables that are
present on a printout from adjustment process. Fields referring to:
- user (max 8 alphanumeric characters)
- project (max 8 alphanumeric characters)
are editable are enable entering a text using balance’s keyboard or connectable external
computer keyboard PS/2 type. The other fields listed in the GLP are set:
- 1 yes (print on a report)
- 0 no (do not print on a report)
GLP submenu is presented in figure 16. If a user applies an external keyboard PS/2 type,
then the relation between the balance’s keyboard and the computer keyboard are as
specified in point 5.1.3 of this user manual.