- 37 -
10.7.4. Coefficient determining process using a set of mass standards
CAUTION:
Before carrying out the process it is necessary to disable AIR BUOYANCY CORRECTION
MODE if it was in use.
Determining air buoyancy correction is also carried out using a dedicated set of 2 pieces of
mass standards. One of the mass standards is made of stainless steel, and the other of
aluminum. Each of the mass standards has specifically determined mass and density.
Determining procedure:
1. Enter the density mode.
Fig. 23-4. Air densitymode selection
2. After entering the mode select appropriate procedure
Fig. 23-5. Air density selecting appropriate mode settings