Send stable measurement result in basic measuring unit
Format: S CR LF
Accessible responses:
- command understood and in progress
-
time limit exceeded while waiting for stable
- command understood but not accessible at this
moment
- command understood and in progress
- response is mass value in basic measuring unit
Mass frame format which is moisture analyzer’s response:
space character mass space unit CR LF
Example:
S CR LF – command sent from a computer
S _ A CR LF - command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response
is mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format: SI CR LF
Accessible responses:
- command understood but not accessible at this
moment
- response is immediate with mass value in basic
weighing unit
Mass frame format which is moisture analyzer’s response:
space unit CR LF
Example:
S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ g _ CR LF - command carried out, immediate
response of mass value in basic weighing unit