Send stable measurement result in current weighing unit
Format: SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
-
time limit exceeded while waiting for stable measurement
result
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF
MASS FRAME
- command understood and in progress
-
response is mass value in current measuring unit
Mass frame format which is moisture analyzer’s response:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U space
stability
marker
space character mass space unit CR LF
Example:
S U CR LFcommand from a computer
S U _ A CR LFcommand understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.
Immediately send measurement result in current measuring unit
Format: SUI CR LF
Accessible responses:
SUI_I CR LF
- command understood but not accessible at this
moment
MASS FORMAT
- mass value in current measuring unit is returned
immediately
Mass frame format which is moisture analyzer’s response:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
Stability
marker
space sign mass space unit CR LF
Example:
S U I CR LFcommand from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ g _ CR LF - command carried out, immediate
response of mass value in current measuring unit
Where: _ - space
Switch on continuous transmission in basic measuring unit
Format: C1 CR LF