n parameter, decimal value specifying working mode number. See OMI
command description for detailed information.
Example
Command:
OMG<CR><LF> give current working mode
Response:
OMG_13_OK<CR><LF> the device currently operates in Statistics mode
Send all implemented commands
Format: PC CR LF
Accessible responses:
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,
LOGIN,LOGOUT,PC" - command carried out, terminal has sent all
implemented commands.
BN Give balance type
Format: BN <CR><LF>
Accessible responses:
BN_A_”x” <CR><LF>
- command understood, balance
type given as a response
BN_I <CR><LF>
- command understood but not
accessible at this moment
xseries of types for a particular balance (in between inverted commas),
with general balance type in front
Example:
Command:
BN <CR><LF> give balance type
Response:
BN_A_”AS” balance type ”AS R”
FS – Give Max capacity
Format: FS <CR><LF>
Accessible responses:
FS_A_”x” <CR><LF>
- command understood, Max
capacity given as a response
FS_I <CR><LF>
- command understood but not
accessible at this moment
xMax value of reading units (in between inverted commas)
Example:
Command:
FS <CR><LF> give Max value for a balance
Response:
FS_A_”220.0000” Max capacity of a balance ”220 g”