EasyManua.ls Logo

RADWAG MA 3Y - Page 147

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 147 -
RV Give program version
Format: RV <CR><LF>
Accessible responses:
RV_A_”x” <CR><LF>
- command understood, program
version given as a response
RV_I <CR><LF>
- command understood but not
accessible at this moment
xprogram version (in between inverted commas)
Example:
Command:
RV <CR><LF> give program number
Response:
RV_A_”r.3.0.9” program version ”r.3.0.9”
A Set AUTOZERO function
Format: A_n <CR><LF>
Accessible responses:
A_OK <CR><LF>
- command carried out
A_E <CR><LF>
- an error occurred on command
carrying out, no parameter
available or incorrect format
A_I <CR><LF>
- command understood but not
accessible at this moment
n parameter, decimal value determining autozero settings
n 0 autozero on
1 autozero off
Caution:
Command changes settings for a current working mode.
Example:
Command:
A_1 <CR><LF> turn autozero function on
Response:
A_OK <CR><LF> autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
EV Set ambient conditions state
Format: EV_n <CR><LF>
Accessible responses:
EV_OK <CR><LF>
- command carried out
EV_E <CR><LF>
- an error occurred on command

Table of Contents

Related product manuals