EasyManuals Logo

RADWAG MA 3Y User Manual

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #148 background imageLoading...
Page #148 background image
- 148 -
carrying out, no parameter
available or incorrect format
EV_I <CR><LF>
- command understood but not
accessible at this moment
n parameter, decimal value determining ambient conditions state
n 0 stable ambient conditions
1 unstable ambient conditions
Caution:
Command changes settings for a current working mode.
Example:
Command:
EV_1 <CR><LF> set value ‘stable’ for ambient conditions option
Response:
EV_OK <CR><LF> ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until
command EV 0 swaps it to value <UNSTABLE>.
FIS – Set filter
FIS (Filtetr, Set)
Command overview:
Command lets the user set filter for a particular device.
Format: FIS_n <CR><LF>
Accessible responses:
FIS_OK <CR><LF>
- command carried out
FIS_E <CR><LF>
- an error occurred on command
carrying out, no parameter
available or incorrect format
FIS_I <CR><LF>
- command understood but not
accessible at this moment
n parameter, decimal value determining filter number
n 1 very fast
2 fast
3 average
4 slow
5 very slow
Caution:
The numbering is assigned to a particular filter name and it is identical for all
balance types.
The command changes settings for a current working mode if, for a particular
balance type, filter settings are assigned to the working mode.