EasyManua.ls Logo

RADWAG MA 3Y - Page 149

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 149 -
Example:
Command:
FIS_1 <CR><LF> set average filter
Response:
FIS_OK <CR><LF> average filter has been set
ARS Set value release
ARS (Approval Result, Set)
Command overview:
Command lets the user set value release options for a particular device.
Format: ARS_n <CR><LF>
Accessible responses:
ARS_OK <CR><LF>
- command carried out
ARS_E <CR><LF>
- an error occurred on command
carrying out, no parameter
available or incorrect format
ARS_I <CR><LF>
- command understood but not
accessible at this moment
n parameter, decimal value determining value release options
n 1 fast
2 fast + reliable
3 reliable
Caution:
The numbering is assigned to a particular value release option and it is
identical for all balance types.
The command changes settings for a current working mode if, for a particular
balance type, value release settings are assigned to the working mode.
Example:
Command:
ARS_1 <CR><LF> set value release parameter to fast+reliable option
Response:
ARS_OK <CR><LF> fast+reliable option has been set
LDS – Set last digit
LDS (Last Digit, Set)
Command overview:
Command lets the user set last digit for a particular device.
Format: LDS_n <CR><LF>
Accessible responses:
LDS_OK <CR><LF>
- command carried out
LDS_E <CR><LF>
- an error occurred on command

Table of Contents

Related product manuals