Example:
Command:
FIS_1 <CR><LF> set average filter
Response:
FIS_OK <CR><LF> average filter has been set
ARS Set value release
ARS (Approval Result, Set)
Command overview:
Command lets the user set value release options for a particular device.
Format: ARS_n <CR><LF>
Accessible responses:
ARS_OK <CR><LF>
- command carried out
ARS_E <CR><LF>
- an error occurred on command
carrying out, no parameter
available or incorrect format
ARS_I <CR><LF>
- command understood but not
accessible at this moment
n parameter, decimal value determining value release options
n 1 fast
2 fast + reliable
3 reliable
Caution:
The numbering is assigned to a particular value release option and it is
identical for all balance types.
The command changes settings for a current working mode if, for a particular
balance type, value release settings are assigned to the working mode.
Example:
Command:
ARS_1 <CR><LF> set value release parameter to fast+reliable option
Response:
ARS_OK <CR><LF> fast+reliable option has been set
LDS – Set last digit
LDS (Last Digit, Set)
Command overview:
Command lets the user set last digit for a particular device.
Format: LDS_n <CR><LF>
Accessible responses:
LDS_OK <CR><LF>
- command carried out
LDS_E <CR><LF>
- an error occurred on command