Page #34 background imageLoading...
Page #34 background image
34
P1.1 iCAL – Internal calibration
start of internal calibration process, process is completely automatic without
interference of balance operator, if pan is loaded display will indicate message
about necessity of removing the load.
or CAL button
P1.2 ECAL – External calibration
calibration with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
balances with internal calibration balances with external calibration
P1.3 tCAL - Test of calibration
comparison of internal calibration mass with its value saved in balance memory.
balances with internal calibration balances with external calibration
P1.4 ACAL – Automatic calibration (balances with internal calibration)
determination of factor, which should decide about starting the automatic internal
calibration
nonE – none of factors will cause start of calibration
tEmP – calibration with regard to change of temperature
timE – calibration with regard to time set in P1.5 CALt
both – calibration with regard to time and temperature.

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
CategoryScales
BrandRADWAG
SeriesPS
CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
DisplayLCD
HousingPlastic
RepeatabilityVaries by model
LinearityVaries by model
CalibrationInternal or external
InterfaceRS232
Power Supply100 - 240V AC
Operating Temperature+15 ÷ +30 °C
Weighing Unitg, lb, oz