EasyManua.ls Logo

Rae RAEGuard 2 PID - Page 33

Rae RAEGuard 2 PID
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAEGuard 2 PID User’s Guide
28
1. Press [-] to move the cursor from “ppm” to “mg/m3” or cycle
through them.
2. Press [+] to register your choice.
3. Press [MODE] to save the selection.
4. Press [+] to save the Measurement Unit, [MODE] to return to
editing it, and [-] to cancel the change.
The new Measurement Unit is saved, and the RAEGuard 2 PID
exits back to the “Monitor Setup” menu.
Meas. Range (Advanced Mode)
Each sensor head has a maximum range for its sensor (for instance,
100 ppm). However, if you want to increase sensitivity in a lower
range, it is possible to change the measurement range (from the preset
100 ppm to 60 ppm, for example).
1. Press [+] to increase the number from 0 to 9 (it wraps around
to 0 again after 9).
2. Press [-] to step through digits.
3. Press [MODE] to save the new value.
If you have set a value that is within range, then you see this screen:
Press [+] to save the new value.
Press [MODE] to go back and edit it again.
Press [-] to exit without making a change.
If the value is within the sensor’s range when you press [+], the value
is saved and “Saved!” is displayed.

Table of Contents

Related product manuals