R&S ESU Instrument Functions
APD Mode
4.96 Operating Manual 1302.6163.12 - 03
Examples of APD
When there is no input signal at the measuring receiver, the envelope of the receiver
noise is shown as in Fig. 4.27 with zero span mode at center frequency fc. The cor-
responding APD response of the receiver noise is shown in Fig. 4.29. The ordinate
axis of the APD function displays the probability that the disturbance envelope
exceeds the disturbance level.
Fig. 4.28 Zero Span of receiver noise
Fig. 4.29 APD of receiver noise
If one impulse is received within 1% of time, as shown in Fig. 4.30, the APD shows
drastic extension to the right side at the bottom of Fig. 4.31, indicating a high distur-
bance level but at low probability. The difference between Fig. 4.28 and Fig. 4.30 is
their peak value (X
max
), it appears dominantly on the abscissa of APD.