Operating Manual 1302.6163.12 - 03 4.257
R&S ESU Instrument Functions
Basic Settings
CENTER B =
MARKER A |
CENTER A =
MARKER B
The CENTER B = MARKER A and CENTER A = MARKER B softkeys couple the
center frequency in diagram B with the frequency of marker 1 in diagram A and the
center frequency in diagram B with the frequency of marker 1 in diagram B. The two
softkeys are mutually exclusive.
This coupling is useful, e.g. for viewing the signal at the marker position in diagram
A with higher frequency resolution or in the time domain in diagram B.
If marker 1 is off, it is switched on and set to the maximum of the trace in the active
diagram.
The softkeys are only available in analyzer mode.
Remote command: INST:COUP CF_B
INST:COUP CF_A
CONFIG
DISPLAY
The CONFIG DISPLAY softkey opens a submenu allowing additional display items
to be added to the screen. In addition, the display power-save mode (DISPLAY
PWR SAVE softkey) and the colors of the display elements can be set here.
SCREEN TITLE The SCREEN TITLE softkey activates the entry of a title for the active diagram A or
B. It switches on or off a title that is already input. The length of the title is limited to
max. 20 characters.
Remote command: DISP:WIND1:TEXT 'Noise Meas'
DISP:WIND1:TEXT:STATe ON
TIME+DATE
ON/OFF
The TIME+DATE ON/OFF softkey switches on or off the display of date and time
above the diagram.
Remote command: DISP:TIME OFF
LOGO ON/OFF The LOGO ON/OFF softkey switches the Rohde & Schwarz company logo dis-
played in the upper left corner of the display screen on or off.
Remote command: DISP:LOGO ON
SCREEN TITLE
TIME+DATE ON/OFF
LOGO ON/OFF
ANNOTATION ON/OFF
DATA ENTRY OPAQUE
DEFAULT COLORS 1/
DEFAULT COLORS 2
DISPLAY PWR SAVE
Side menu
SELECT OBJECT
BRIGHTNESS
TINT
SATURATION
PREDEFINED COLORS