Operating Manual 1302.6163.12 - 03 4.271
R&S ESU Instrument Functions
Basic Settings
TRANSDUCER
FACTOR
The TRANSDUCER FACTOR softkey places the scrollbar on the position of the
active transducer factor.
If a transducer factor is not active, the scrollbar is placed on the first line of the table.
Remote command: CORR:TRAN:SEL <name>
CORR:TRAN ON | OFF
TRANSDUCER
SET
The TRANSDUCER SET softkey places the scrollbar on the position of the active
transducer set. For further details refer to “Entry and Editing of Transducer Sets” on
page 4.275.
If no transducer set is switched on, the bar is set to the first line of the table.
Remote command: CORR:TSET:SEL <name>
CORR:TSET ON |OFF
EDIT The EDIT TRD FACTOR softkey gives access to the submenu for editing and gener-
ating transducer factors. For details refer to section “Entry and Editing of Transducer
Factors” on page 4.272.
NEW The NEW FACTOR softkey gives access to the submenu for editing and generating
transducer factors. For details refer to section “Entry and Editing of Transducer Fac-
tors” on page 4.272.
DELETE The DELETE softkey deletes the marked factor or set.
To prevent deletion by mistake, deletion has to be confirmed.
Remote command: CORR:TRAN DEL
VIEW
TRANSDUCER
The VIEW TRANSDUCER softkey activates the display of the active transducer fac-
tor or set. During the measurement, the interpolated characteristic of the active
transducer factor or set is shown in the diagram instead of the measured values.
Remote command: CORR:TRAN:VIEW ON
VIEW
TRANSDUCER
The VIEW TRANSDUCER softkey activates the display of the active transducer fac-
tor or set. During the measurement, the interpolated characteristic of the active
transducer factor or set is shown in the diagram instead of the measured values.
Remote command: CORR:TRAN:VIEW ON
REFLVL ADJ
AUTO MAN
When a transducer factor is used, the trace is moved by a calculated shift. However,
an upward shift reduces the dynamic range for the displayed values. With the soft-
key REFLVL ADJ an automatic reference level offset adaptation allows to restore
the original dynamic range by also shifting the reference level by the maximum
value of the transducer factor.