Operating Manual 1302.6163.12 - 03 4.7
R&S ESU Instrument Functions
R&S ESU Initial Configuration – PRESET Key
Table 4-1 RECEIVER SETUP of R&S ESU
Parameter Settings
Mode IF (bar graph in screen A and IF spectrum in screen B)
Center frequency step size AUTO COARSE
RF attenuation auto (10 dB minimum)
Level range 100 dB log
Level unit dBµV
Sweep time auto (2,5 ms)
Resolution bandwidth Bargraph 120 kHz / 6 dB, IF bandwidth 10 kHz
FFT filters off
Sweep cont
Trigger free run
Trace 1 clr write, autopeak
Cal correction on
Noise source off
Input RF 1, AC-coupled
Display Split screen
Tracking generator (only with
option R&S FSP-B9)
off
External generator 1/2 (only
with option R&S FSP-B10)
off
Preamplifier on
Preselector off