R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
6.12 Operating Manual 1302.6163.12 - 03
6.5 CALCulate Subsystem
The CALCulate subsystem contains commands for converting instrument data, transforming and carry-
ing out corrections. These functions are carried out subsequent to data acquisition, i.e. following the
SENSe subsystem.
The numeric suffix is used in CALCulate to make the distinction between the two measurement windows
SCREEN A and SCREEN B:
CALCulate1 = Screen A
CALCulate2 = Screen B.
For commands without suffix, screen A is selected automatically.
Full Screen The settings are valid for the measurement window selected with the numeric suffix.
They become effective as soon as the corresponding measurement window has been
selected as active measurement window using the command DISPLay[:WIN-
Dow<1|2>]:SELect. Triggering measurements and querying measured values is
possible only in the active measurement window.
Split Screen The settings are valid for the measurement window selected by means of the numeric
suffix and become effective immediately.
In receiver mode, the marker can be activated only for the scan display (numeric
suffix 2) and after a scan is carried out.