Operating Manual 1302.6163.12 - 03 6.49
R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:MINimum:AUTO ON | OFF
This command switches an automatic minimum peak search for marker 1 at the end of each particular
sweep on and off. The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account. The numeric suffix at MARKer<1...4> is irrelevant.
Example: "CALC:MARK:MIN:AUTO ON"
Activates the auto search function for marker 1.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
CALCulate<1|2>:MARKer<1...4>:MINimum[:PEAK]
This command positions the specified marker to the current minimum value of the corresponding trace.
The corresponding marker is activated first or switched to marker mode, if necessary.
Example: "CALC:MARK2:MIN"
'Positions marker 2 in screen A to the minimum value of the trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MINimum:NEXT
This command positions the specified marker to the next higher minimum value of the corresponding
trace.
Example: "CALC:MARK2:MIN:NEXT"
'Positions marker 2 in screen A to the next higher maximum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value and no query.
If no minimum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: -200) is produced.
If no next higher minimum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.