Operating Manual 1302.6163.12 - 03 6.97
R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
6.5.9 CALCulate:UNIT Subsystem
The CALCulate:Unit subsystem defines the units for power measurement settings.
CALCulate<1|2>:UNIT:POWer DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT | DBUA |
AMPere | DBPT | DBUV_M | DBUA_M
This command selects the unit for power in the selected measurement window.
In receiver mode, only the units DBUV, DBUV_M, DBUA, DBUA_M, DBPW, DBPT and DBMV are
available.
Example: "CALC:UNIT:POW DBM"
'Sets the power unit for screen A to dBm.
Characteristics: *RST value: dBm
SCPI: device-specific
Mode: R, A