R&S ESU Remote Control – Description of Commands
DIAGnostic Subsystem
6.102 Operating Manual 1302.6163.12 - 03
DIAGnostic<1|2>:SERVice:NSOurce ON | OFF
This command switches the 28 V supply of the noise source at the rear panel on or off.
The numeric suffix <1|2> is ignored with this command.
Example: "DIAG:SERV:NSO ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A
DIAGnostic<1|2>:SERVice:CSOource[:POWer] <numeric_value>
This command switches the level of the 128 MHz reference signal source between 0 dBm and -30 dBm
in the selected measurement window.
Example: "DIAG:SERV:CSO 0DBM"
Characteristics: *RST value: -30 dBm
SCPI: device-specific
Mode: R, A
DIAGnostic<1|2>:SERVice:STESt:RESult?
This command reads the results of the selftest out of the instrument. The lines of the result table are
output as string data separated by commas:
The numeric suffix <1|2> is ignored with this command.
Parameter: "Total Selftest Status: PASSED","Date (dd/mm/yyyy): 09/07/
2006
TIME: 16:24:54","Runtime: 00:06", ...
Example: "DIAG:SERV:STES:RES?"
Characteristics: *RST value: --
SCPI: device-specific
Mode: R, A
This command is only a query and therefore has no *RST value.