Operating Manual 1302.6163.12 - 03 6.131
R&S ESU Remote Control – Description of Commands
INPut Subsystem
INPut<1|2>:LISN[:TYPE] TWOPhase | FOURphase | ESH3Z5 | ESH2Z5 | ENV4200 | ENV216 | OFF
This command selects the V-network that is controlled via the USER port. In receiver mode, this setting
is used also for the preliminary measurement.
Example: ”:INP:LISN:TWOP"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A
INPut<1|2>:LISN:PHASe L1 | L2 | L3 | N
This command selects the phase of the V-network that is used and which is controlled via the USER
port. In receiver mode, this setting is used also for the preliminary measurement. The permissible
selection depends on the selected V-network.
Example: ":INP:LISN:PHAS L1"
Characteristics: *RST value: L1
SCPI: device-specific
Mode: R, A
INPut<1|2>:LISN:PEARth GROunded | FLOating
This command selects the setting of the Protective EARth that is controlled via the USER port. In
receiver mode, this setting is used also for the preliminary measurement. The availability depends on
the selected V-network.
Example: ":INP:LISN:PEAR GRO"
Characteristics: *RST value: GROunded
SCPI: device-specific
Mode: R, A
INPut<1|2>:LISN:FILTer:HPAS[:STATe] ON | OFF
This command selects the setting for the highpass on the V-network that is controlled via the USER
port. In receiver mode, this setting is used also for the preliminary measurement. This command is
available only for the R&S ENV 216 V-network (INP:LISN ENV216).
Example: ":INP:LISN:FILT:HPAS ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R
TWOPhase and ESH3Z5 R&S ESH3-Z5 (two phases and protective earth are controllable)
FOURphase and ESH2Z5 R&S ESH2-Z5 (four phases and protective earth are controllable)
ENV4200 R&S ENV 4200 (four phases are controllable)
ENV216 R&S ENV 216 (two phases and highpass are controllable)
OFF Remote control deactivated