R&S ESU Remote Control – Description of Commands
INPut Subsystem
6.132 Operating Manual 1302.6163.12 - 03
INPut<1|2>:IMPedance 50 | 75
This command sets the nominal input impedance of the instrument. The set impedance is taken into
account in all level indications of results.
The setting 75 should be selected, if the 50 input impedance is transformed to a higher impedance
using a 75 adapter of the RAZ type (= 25 in series to the input impedance of the instrument). The
correction value in this case is 1.76 dB = 10 log (75/ 50).
Example: "INP:IMP 75"
Characteristics: *RST value: 50
SCPI: conform
Mode: R, A
INPut<1|2>:MIXer[:POWer] <numeric value>
This command defines the desired power at the input mixer of the R&S ESU. On any change to the
reference level the RF attenuation will be adjusted in a way that makes the difference between
reference level and RF attenuation come as close to the desired mixer level as possible.
Example: "INP:MIX -30"
Characteristics: *RST value: - 25 dBm
SCPI: device-specific
Mode: A
INPut<1|2>:MIXer:AUTO ON | OFF
This command enables/disables the automatic setup of the mixer level.
Example: "INP:MIX:AUTO ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: A
INPut<1|2>:PRESelection[:STATe] ON | OFF
In analyzer mode, this command switches the preselection on or off.
Example: ":INP:PRES:STAT ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
In receiver mode, the preselection is always switched on and cannot be switched
off.