Operating Manual 1302.6163.12 - 03 6.165
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]CORRection:TSET[:STATe] ON | OFF
This command switches the selected transducer set on or off.
Example: "CORR:TSET:SEL 'SET1'"
"CORR:TSET ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A
[SENSe<1|2>:]CORRection:TSET:DELete
This command deletes the selected transducer set.
Example: "CORR:TSET:SEL 'SET1'"
"CORR:TSET:DEL"
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value.
[SENSe<1|2>:]CORRection:YIG:TEMPerature:AUTO ON | OFF
This command switches on or off the automatic correction of the YIG filter frequency drift.
When correction is switched on, it is checked once per minute whether the temperature on the YIG
filter has changed by more than 5K relative to the last instance of correction. If this is the case, the
frequency of the YIG filter is – at the end of the next measurement – adjusted as required for the new
temperature. For time-critical applications, the correction function can be switched off after an
operating period of 30 minutes.
Example: "CORR:YIG:TEMP OFF"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: R, A