EasyManua.ls Logo

R&S ESU Series - Page 605

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
6.166 Operating Manual 1302.6163.12 - 03
6.16.4 SENSe:DEMod Subsystem
The SENSe:DEMod subsystem controls the analog demodulation of the video signal. The measurement
window is selected by SENSe1 (screen A) and SENSe2 (screen B).
[SENSe<1|2>:]DEMod OFF | AM | FM
This command selects the type of analog demodulation.
Example: "DEM FM"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R
[SENSe<1|2>:]DEMod:SQUelch:LEVel <numeric_value> | MIN | MAX | DEF
This command enables the input of a level threshold below which the audible AF is cut off.
Example: "SENSe:DEM:SQU:LEV 80"
'Sets the squelch level to 80.
Characteristics: *RST value: 60
SCPI: device-specific
Mode: R