R&S ESU Remote Control – Description of Commands
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E
6.282 Operating Manual 1302.6163.12 - 03
6.23.4 856x: Emulation of the Spurious Response Measurement Utility
85672A
6.23.4.1 General commands for Spurious
6.23.4.2 Commands for TOI-Measurement
Command Supported subset Function Corresponding
HP-Models
Status
SP_OK A value of 1 denotes a
successful measurement
HP 856xE available in V4.1x and
above
SP_RMT A value of 1 means, that the
instrument settings are
saved to register 9
HP 856xE available in V4.1x and
above
SP_EXTREF
0: use internal reference;
1. use external reference
HP 856xE available in V4.1x and
above
SP_EXIT Quits spurious HP 856xE available in V4.1x and
above
Command Supported subset Function Corresponding
HP-Models
Status
SP_TOI Executes TOI
measurement
HP 856xE available in V4.1x and
above
SP_TOIFA Lower primary signal
frequency in Hz
HP 856xE available in V4.1x and
above
SP_TOIFB Upper primary signal
frequency in Hz
HP 856xE available in V4.1x and
above
SP_TOIFS Primary signal frequency
spacing in Hz
HP 856xE available in V4.1x and
above
SP_TOI_SL Lower signal amplitude in
dBm
HP 856xE available in V4.1x and
above
SP_TOI_SU Upper signal amplitude in
dBm
HP 856xE available in V4.1x and
above
SP_TOI_PL Lower distortion product
amplitude in dBm
HP 856xE available in V4.1x and
above
SP_TOI_PU Upper distortion product
amplitude in dBm
HP 856xE available in V4.1x and
above
SP_TOI_A,
SP_TOI_B
Third or intercept point in
dBm
HP 856xE available in V4.1x and
above
SP_TOINA,
SP_TOINB
The flag=0 indicates a good
measurement
HP 856xE available in V4.1x and
above