Operating Manual 1302.6163.12 - 03 6.329
R&S ESU Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
R&S
ESU
FSE [SENSe<1|2>:]SWEep:TIME 2.5ms to 1000s | 1µs to
16000s
different parameter
ranges for R&S ESU and
FSE
R&S
ESU
FSE [SENSe<1|2>:]SWEep:TIME:AUTO ON | OFF
FSE [SENSe<1|2>:]TCAPture:LENGth 1024 | 2048 | 4096 |
8192 | 16384
not available for
R&S ESU
FSE [SENSe<1|2>:]TV:PSOFfset 0 to 6.5 MHz not available for
R&S ESU and FSET
R&S
ESU
[SENSe<1|2>:]TV:CCVS INTernal | EXTernal new function for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]TV[:STATe] ON | OFF
R&S
ESU
FSE SOURce:AM:STATe ON | OFF
R&S
ESU
FSE SOURce:DM:STATe ON | OFF
R&S
ESU
SOURce:EXTernal[:STATe ] ON | OFF new command for
R&S ESU
R&S
ESU
SOURce:EXTernal:FREQuency:OFFset <numeric_value> new command for
R&S ESU
R&S
ESU
SOURce:EXTernal:FREQuency[:FACTor]:
NUMerator
<numeric_value> new command for
R&S ESU
R&S
ESU
SOURce:EXTernal:FREQuency[:FACTor]:
DENominator
<numeric_value> new command for
R&S ESU
R&S
ESU
SOURce:EXTernal:FREQuency:SWEep[:STATe] ON | OFF new command for
R&S ESU
R&S
ESU
SOURce:EXTernal:POWer[:LEVel] <numeric_value> new command for
R&S ESU
R&S
ESU
FSE SOURce:FM:STATe ON | OFF
R&S
ESU
FSE SOURce:FREQuency:OFFSet -150Hz to 150MHz different value ranges for
R&S ESU and FSE
FSE SOURce:POWer:ALC:SOURce INTernal | EXTernal not available for
R&S ESU and FSET
R&S
ESU
FSE SOURce:POWer[:LEVel][:IMMediate]:OFFSet -200dB to +200dB
R&S
ESU
FSE SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] -30dBm to 0dBm different value ranges for
R&S ESU and FSE
R&S
ESU
FSE STATus:OPERation:CONDition?
FSE STATus:OPERation:CONDition?
R&S
ESU
FSE STATus:OPERation:ENABle 0 to 65535
R&S
ESU
FSE STATus:OPERation:NTRansition 0 to 65535
Devices Command (Sheet 41 of 47) Parameter Notes