Operating Manual 1302.6163.12 - 03 9.5
R&S ESU Error Messages
SCPI-Specific Error Messages
-158 String data not allowed
The command contains a valid string at a position which is not allowed.
Example: A text parameter is set in quotation marks, INPut:COUPling "DC"
-160 Block data error
The command contains faulty block data.
-161 Invalid block data
The command contains faulty block data.
Example: An END message was received prior to reception of the expected number
of data.
-168 Block data not allowed
The command contains valid block data at an invalid position.
Example: The command *RCL requires a number to be indicated.
-170 Expression error
The command contains an invalid mathematical expression.
-171 Invalid expression
The command contains an invalid mathematical expression.
Example: The expression contains mismatching parentheses.
-178 Expression data not allowed
The command contains a mathematical expression at an invalid position.
Error code Error text in the case of queue poll
Error explanation
-200 Execution error
Error on execution of the command.
-201 Invalid while in local
The command is not executable while the device is in local due to a hard local control.
Example: The device receives a command which would change the rotary knob state,
but the device is in local so the command can not be executed.
-202 Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to
LOCS from REMS or to LWLS from RWLS.
-210 Trigger error
Error on triggering the device.
-211 Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored because of device timing
considerations.
Example: The device was not ready to respond.
-212 Arm ignored
An arming signal was ignored by the device.
-213 Init ignored
Measurement initialization was ignored as another measurement was already in
progress.
Error code Error text in the case of queue poll
Error explanation