R&S ESU Error Messages
Device-Specific Messages
9.8 Operating Manual 1302.6163.12 - 03
9.3 Device-Specific Messages
Table 9-1 Device-specific messages
Error code Error text in the case of queue poll
Error explanation
-400 Query error
General error occurring when data are requested by a query.
-410 Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has
been sent completely.
-420 Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430 Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue
operation.
-440 Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite
response.
Error code Error text in the case of queue poll
Error explanation
1036 MS: The correction table based amplifier gain exceeds the amplifier range for
CALAMP1 and CALAMP2 on IF board
This message is output when the setting range of the calibration amplifier is not
sufficient for the required correction. The error occurs only with modules which are not
correctly adjusted or defective.
1052 Frontend LO is Unlocked
This message is output when the phase regulation of the local oscillator fails in the RF
frontend.
1060 Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is output when the gate signal length is not sufficient for the pull-in delay
with a predefined gate delay.
2022 OPTIONS.INI invalid
This message is output when an error has been recognized in the file OPTIONS.INI
which contains the clearing codes for retrofitable firmware applications. If this file is not
correctly recognized, all firmware applications are blocked for this instrument.
2028 Hardcopy not possible during measurement sequence
This message is output when a printout is started during scan sequences that cannot
be interrupted. Such sequences are for example:
Recording the system error correction data (calibration)
Instrument selftest
In such cases a synchronization to the end of the scan sequence should be performed
prior to starting the printout.