Operating Manual 1302.6163.12 - 03 9.7
R&S ESU Error Messages
SCPI-Specific Error Messages
-253 Corrupt media
The media is corrupt.
Example: The memory stick has a defect.
-254 Media full
The media is full.
Example: There is no space on the memory stick.
-255 Directory full
The media directory is full.
-256 File name not found
The file name cannot be found on the media.
-257 File name error
The file name is wrong.
Example: An attempt is made to copy to a duplicate file name.
-258 Media protected
The media is protected.
Example: The write-protect tab on the memory stick is activated.
-260 Expression error
The expression contains an error.
Error code Error text in the case of queue poll
Error explanation
-300 Device-specific error
R&S ESU-specific error not defined in greater detail.
-310 System error
This error message suggests an error within the instrument. Please inform the R&S
Service.
-313 Calibration memory lost
Loss of the non-volatile data stored using the *CAL? command. This error occurs
when the correction data recording has failed.
-330 Selftest failed
The selftest could not be executed.
-350 Queue overflow
This error code is entered in the queue instead of the actual error code if the queue is
full. It indicates that an error has occurred but not been accepted. The queue can
accept 5 entries.
Error code Error text in the case of queue poll
Error explanation