LT  ± 1% of reading or ± 1°C for Tmeas > 0°C (32°F) 
(whichever is greater) 
MT  ± 1% of reading for Tmeas > 350°C (662°F) 
± 2°C or ± 2% for Tmeas < 350°C (662°F) 
(whichever is greater) 
G5L, G5H, G7  ± 1% of reading 
3M  ± 1% of reading for Tmeas > 150°C (302°F) 
  ± 5°C for Tmeas < 150°C (302°F) 
2ML  ± (0.3% of reading + 2°C) 
2MH  ± (0.3% of reading + 1°C) 
1ML  ± (0.3% of reading + 1°C) for Tmeas > 450°C (842°F) 
± (2% of reading + 2°C) for Tmeas < 450°C (842°F) 
1MH  ± (0.3% of reading + 1°C) for Tmeas > 650°C (1202°F) 
± (2% of reading + 2°C) for Tmeas < 650°C (1202°F) 
 
Repeatability
 
LT, MT, G5, 3M, G7  ± 0.5% of reading or ± 0.5°C, whichever is greater 
2ML, 2MH  ± (0.1% of reading + 1°C) 
1ML  ± (0.1% of reading + 1°C) for Tmeas > 450°C (842°F) 
± (1% of reading + 1°C) for Tmeas < 450°C (842°F) 
1MH  ± (0.1% of reading + 1°C) for Tmeas > 650°C (1202°F) 
± (1% of reading + 1°C) for Tmeas < 650°C (1202°F) 
 
Temperature Resolution (mA output) 
2MH, 1MH  0.2 K 
all other models  0.1 K 
Noise Equivalent Temperature (NETD) 
LT  0.1 K at Tobj = 23°C (73°F), Tamb = 23°C (74°F) 
3M  0.1 K at Tobj = 200°C (392°F), Tamb = 25°C (77°F) 
MT, G5, G7  0.5 K at Tobj = 10% of full measurement range, Tamb = 25°C (77°F) 
2M, 1M  0.5 K at Tobj = 10% of full measurement range, Tamb = 25°C (77°F) 
Response time = Instrument Response time 
Emissivity  0.100 to 1.150, in 0.001 increments 
Signal Processing  Peak hold, valley hold, averaging, advanced peak hold, 
advanced valley hold, ambient background temperature 
compensation