Accessories 
Marathon MM  Rev. D7 Jul 2017  63 
9.10 Low Temp Cable 
The  12-wire  low  temp cable  (XXX2CLTCB…)  is  used for  wiring  the  sensor  with the  24 VDC  power 
supply, all outputs, and the RS485 interface. The cable is PUR (Polyurethane) coated and withstands 
ambient temperatures form -40 to 105°C (-40°F to 221°F). PUR coated cables are flexible and have good 
to excellent resistance to against oil, bases, and acids. 
  Temperature:  -40 to 105°C (-40°F to 221°F) 
  Cable material  PUR- 11Y (Polyurethane), Halogen free, Silicone free 
  Cable diameter:  7.2 mm (0.283 in) nominal 
  Conductors: 
Power supply  2 wires (black/red) 
Conductor:  0.2 mm² (AWG 24), 7x32 tinned copper 
Isolation:  PE- 2YI1 
Shield:  none 
RS485 interface  2 twisted pairs (black/white and purple/gray) 
Conductor:  0,2 mm² (AWG 24), 7x32 tinned copper 
Isolation:  PE- 2YI1 
Shield:  CDV-15, 85% covered 
Outputs and Ground  6 wires (green/brown/blue/orange/yellow/clear) 
Conductor:  0,2 mm² (AWG 24), 7x32 tinned copper 
Isolation:  PE- 2YI1 
Shield:  none 
Further information for wiring the cable can be found in section 6.2 Electrical Installation, page 28. 
The  low  temp  cable  can  be  purchased  from  the  manufacturer  in  the  following  lengths: 
4 m, 8 m, 15 m, 30 m, 60 m (13 ft., 26 ft., 49 ft., 98 ft., 197 ft.) 
If you cut the cable to shorten it, notice that both sets of twisted-pair wires have drain wires inside their 
insulation. These drain wires (and the white wire that is not part of the twisted pair) must be connected 
to the terminal labeled CLEAR. 
If you  purchase your  own  RS485 cable,  use  wire  with the  same  specifications as  those  listed above. 
Maximum RS485 cable length is 1.200 meters (4000 ft).