EasyManua.ls Logo

Reichert AT550 - Flyer Readings Explained

Reichert AT550
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Instructions for Use - Low Condence Readings
Instructions for Use - Low Condence Readings
AT550 measurements are made in a few milliseconds and occur at random relative to
the ocular pulse. Normal measurement uctuation of 2 to 4 mmhg should be expected
due to the cardiac-related pulse amplitude. Occasionally an erroneously high reading
may appear (a “yer”). The AT550 distinguishes these readings by putting brackets
around the value, as shown on the screen below.
A “yer is used to compute the average, which would create an erroneously high
value. By taking a fourth measurement, the AT550 will automatically replace the
yer” value.

Other manuals for Reichert AT550

Related product manuals