38
Fuses
Fuses are located above the power inlet. (See p. 5, No. 10 for location.) Only replace
fuses with T 0.315 AL 250V @ 230V and T 0.63 AL 250V @ 115V as described on the
power inlet panel.
Internal fuse (F1) located on power I/0 board replace with T6.3 AL 250V model
number 13900 and 13901. Replacement to be performed by qualied service
personnel only.
Forehead Rest
For hygienic reasons, after each patient, wipe the forehead rest with a clean cloth
or with alcohol wipes.
PC Card
The PC card slot is located behind the printer door. Before removing or installing
a PC card, disconnect the AT550 from the power supply and follow installation
Maintenance