Step 3: Mix control solution by gently
inverting the bottle. Remove the cap from
the control solution bottle. While the
Apply Blood symbol isashing,place
a very small drop on the end of the
test strip. The test strip will draw up the
solution and the meter will beep. DO NOT
put control solution on top of the test
strip. The meter will begin to count down.
Step 4: In 7 seconds, the meter will
beep again and display the control
solution result. Compare the result to
the range printed on the test strip
bottle. Make sure the result is within the
acceptable range. If the result falls within
this range, the meter and test strip are
working correctly.