OPTICAL ON
OPTICAL OFF
OPTICAL ON
TIMER OFF
OPTICAL ON
OPTICAL OFF
OPTICAL ON
TIMER OFF
CONTINUOUS USE
5% USAGE - 72 min/day
STAND-BY LIFE
Minimum Minimum Minimum Minimum Minimum
98 days365 days 80 days 140 hrs 110 hrs
BATTERY LIFE EXPECTANCY
Alkaline battery
Duracell type MN 1604 or equivalent
Typical battery reserve life.
Using an alkaline battery at 5% usage,
the probe will typically continue to
operate for 8 hours, after the
MI 12/OMI low battery LED lights up.
Dead battery indication
When the battery voltage drops below
the threshold where performance can
be guaranteed, the MP10 probe status
LED will change to constant red.
The probe output relay will also be
forced into its open state, causing the
machine to stop, until a new battery is
inserted.
The probe will revert to the stand-by
mode after changing the battery.
1-14
Probe battery
Power for the probe is supplied by a type PP3
9V battery.
The Probe status LED indicates when the
battery has come to the end of its useful life.
Low battery indication
When MI 12 or OMI low battery LED lights up,
battery voltage is low and the end of useable
battery life is approaching.
(The low battery LED will only light up during
the probe operating mode)
- see APPENDIX pages 1-39 and 1-40.
The machine control may also be programmed
to flag up a low battery alarm.