RAMP® System Operation
Operator’s Manual, RAMP
®
Reader 24
Use [▲] or [▼] and [Enter] to select the desired option. If YES is selected, the following screen will
be displayed:
Upon completion, the screen returns to the most recent sample result.
After the last result is deleted from the VIEW menu, the following screen is displayed:
Upon completion the screen returns to the OPTIONS Menu.
2.5.3 Run Internal Quality Control (IQC)
IQC can be scheduled to run automatically at intervals during a 24-hour period (Section 2.6.9). In
addition, a manual IQC test can be performed at any time by performing the following steps.
From the RUN/OPTIONS Menu, select Options and press [▲] or [▼] until 3. RUN INTERNAL QC is
displayed. Press [Enter] to select. The test takes just a few seconds, during which the following
screen is displayed:
Upon completion of IQC, the screen returns to the OPTIONS Menu.
If an error occurs, refer to Section 4.1.
NOTE: A manual IQC does not affect the IQC timer settings.
CONDUCTING IQC
PLEASE WAIT
LAST RESULT
DELETED!
RESULT HAS BEEN
DELETED!