Fault messages
40
9 Fault messages
Flashing sequence
Meaning
Instruction
G4 and G5 flash
alternately
Safety circuit is not
closed
Close the door.
Pull the locking bracket forwards.
Allow the locking bolt to engage.
G5 flashes
Motor block
Remove the blocking pieces of sample material
from the grinding chamber.
Press button G6
Press button G1.
G4 + G5 flash
simultaneously
Faulty relay contact
Service necessary!
G4 flashes
Revolving field
detection
Switch the two phases (L1 with L2) on the power
cable
Fig. 32:Flashing sequence during error message
G1
G4
G5
G6