EasyManua.ls Logo

RH CGS-240 - Page 45

RH CGS-240
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
CGS-240 Manual 6/15/2021
Page 45 of 148
2) Send the pressure sensors to a lab that has the capability to calibrate them independent of
the CGS-240.
a. The lab must be able to power them and communicate with them over RS-485.
b. The lab must be able to perform absolute pressure calibration, read data via RS-485,
compute calibration coefficients, and store these calibration coefficients in the
transducer’s memory.
Sensor Parameters: The sensors are Mensor CPT6020, with a required supply voltage
of 12VDC, and communicating via RS-485 at 9600/8-N-1. Sensor P
L
is on Address 0, P
H
is
on Address 1, and the units are in Pascal (Pa).
IMPORTANT: If the lab changes any of the parameters during calibration, the sensors
should be returned to the above sensor parameters by the calibration lab prior to
return and subsequent re-installation.
3) Upon return of the fully calibrated sensors, the CGS-240 must have
Default Coefficients entered. The use of default coefficients ensures
that the instrument will not perform any post-measurement
corrections to the newly calibrate pressure sensor data.
Coefficient
Default
a0
0
a1
1
a2
0
a3
0
4) Select the Settings icon, then select [Pl] Low
Pressure Coefficients.
5) Enter the default coefficients a0, a1, a2 and a3 (as
shown), then click the Save button.
Repeat steps 5 and 6 for [PH] High Pressure Coefficient.
Refer to Section 5.1.3 Pressure Sensor Reinstallation.