Chapter 6 Sweep RIGOL
DG4000 Series Users Guide
6-11
3. Manual Trigger
A sweep will be generated from the corresponding channel once you press
Trigger1 or Trigger2 at the front panel.
Trigger Output Edge
In Sweep mode, when Int or Manual trigger source is selected, the generator
will output a TTL compatible signal with specified edge from the [Mod/FSK/Trig]
connector at the rear panel.
Internal Trigger: the generator outputs a Square waveform with 50% duty
cycle from the [Mod/FSK/Trig] connector at the beginning of the Sweep.
The trigger period depends on the specified sweep time, return time, start
hold and end hold.
Manual Trigger: the generator outputs a Pulse waveform with greater than
1μs pulse width at the beginning of the Sweep from the [Mod/FSK/Trig]
connector.
External Trigger: the [Mod/FSK/Trig] connector is used as the input
terminal of external trigger signal and has no trigger output.