Chapter 13 Specifications RIGOL
DG4000 Series User’s Guide
Chapter 13 Specifications
All the specifications can be guaranteed if the following two conditions are met
unless where noted.
The generator is within the calibration period and has performed
self-calibration.
The generator has been working continuously for at least 30 minutes under
the specified temperature (18°C ~ 28°C).
All the specifications are guaranteed unless those marked with “typical”.
Sine, Square, Ramp, Pulse, Noise, Harmonics
150 kinds, including Sinc, Exponential Rise, Exponential Fall, ECG,
Gauss, HaverSine, Lorentz, Dual-Tone, DC, etc.
Frequency Characteristics
Sine Wave Spectrum Purity
Typical (0dBm)
DC-1MHz: <-60dBc