RIGOL Chapter 1 Specifications
DP1308A Service Guide
1-2
Voltage
0.03%+1mV
0.02%+2mV
Current
0.1%+3mA
0.05%+1mA
Voltage Programming Speed (within 1% of total variation range)
Rising
Full
Load
11ms
50ms
No Load
10ms
45ms
Falling
Full
Load
13ms
20ms
No Load
200ms
400ms
OVP/OCP
Accuracy ± (output
percentage + offset)
0.5%+0.5V/0.5%+0.5A
Activation Time
1.5ms (OVP 3V)
<10ms (OVP <3V and OCP)
Machenical
Dimension
235 mm (W) x 155 mm (H) x 384 mm (D)
Weight
8.5 kg
Power Supply
AC Output
(50Hz to 60Hz)
100Vac+10%, 115Vac+10%, 220Vac+10%,
230Vac+10% (250VAC Max)
Environment
Working Temperature
Full rated output: 0 to 40
At higher temperature: the linearity of the output
current falls to 50% at the maximum temperature 55
Cooling Method
Fan cooling
Remarks:
[1] Specifications are for one hour warm-up at 25.
[2] The maximum time required for regulating corresponding output after receiving the APPLy and
SOURce commands.
[3] The variation of the output within 8 hours after the instrument is warmed up for 30 minutes
and both the load circuit and environment temperatures are constant.