Chapter 5 To Trigger the Oscilloscope RIGOL
DS1000Z Users Guide
5-35
--press Data Bits to select “5 bits, “6 bits, “7 bits” or “8 bits”;
--press Data and input the data value according to the setting in Data Bits and
the upper limits are 31, 63, 127 and 255 respectively.
Baud Rate:
Set the baud rate of data transmmision (equal to specifying a clock frequency). Press
Baud to set the desired baud rate to 2400 (default), 4800, 9600, 19200, 38400,
57600, 115200 and user-defined, the default unit is bps. When “User” is selected,
press Setup and use to set a more specific value from 1 to 900000 with an
adjustment step of 1 bps.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameter (noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of
Trigger Levelon page 5-9.

Table of Contents

Other manuals for Rigol DS1000Z Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000Z Series and is the answer not in the manual?

Rigol DS1000Z Series Specifications

General IconGeneral
BrandRigol
ModelDS1000Z Series
CategoryTest Equipment
LanguageEnglish

Related product manuals