RIGOL Chapter 15 System Function Setting
MSO2000A/DS2000A Users Guide
15-12
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5 . Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channel connections, and then press Utility Self-Cal to
open the self-calibration menu and self-calibration interface.
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note: Most of the keys are disabled during the self-calibration.