2. Important Safety Information 2-1. Danger information
10 / 79
The battery specifications are as follows:
<Dry cell specifications>
The explosion-proof class is Ex ia IIC T4 Ga.
The ratings are as follows:
Japan models:
Power source: 3 V DC (Toshiba LR03 battery × 2)
Ambient temperature: 40 °C to +60 °C
Export models:
Power source: 3 V DC, 1 mA (Toshiba LR03, Duracell MN2400, or Duracell PC2400 battery × 2)
Ambient temperature: 40 °C to +60 °C
Does not accept rechargeable batteries.
<Rechargeable battery specifications>
The explosion-proof class is Ex ia IIC T3 Ga.
The ratings are as follows:
Power source: 2.4 V DC, 1 mA (Panasonic eneloop (BK-4MCC) battery × 2)
Ambient temperature: 40 °C to +60 °C
Use two eneloop (BK-4MCC) (Panasonic) batteries. Does not accept dry cell batteries.
Use a BQ-CC23 (Panasonic, Delta Vt control recharging) recharger.
The rating for recharging is 1.5 V DC, 550 mA.
Recharge the batteries only in nonhazardous locations.
If the product is used as an explosion-proof device, note that the explosion-proofing rating conditions specify the
battery type to be used.
The battery types are as follows:
<Dry cell specifications>
Japan models:
Power source: 3 V DC, 1 mA (Toshiba LR03 battery × 2)