Rinstrum - 5200 Communications Manual Rev 1.2
Page 10 0025-605-120
2.4. Command Details
2.4.1. ACL: Set Automatic Temperature Calibration On/Off
This is used to set the Automatic Temperature Calibration On/Off.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
2 - - - - - - Immediately No
Parameter Details
Parameter Description Range Default Required
1 Temperature
calibration during no
motion
0 to 1 (0: Off, 1: On) 1 No
2 Temperature
calibration during
motion
0 to 1 (0: Off, 1: On) 1 No
The 5200 performs an automatic temperature calibration to ensure that changes in
ambient temperature do not affect accuracy. During this calibration no new data is
received. The time taken for this calibration depends on the sync frequency
according to the following formula: 6 / sync frequency (seconds).
This feature cannot be used in Trade mode and both parameters are set to ON
when the instrument is reset.