Rinstrum - 5200 Communications Manual Rev 1.2
Page 54 0025-605-120
2.4.42. WMD: Set Weighing Mode
Set the weighting mode of the instrument. This selects between Single Range, Dual
Range and Dual Interval weighing modes.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
2 With TDD1 Immediately Yes
Parameter Details
Parameter Description Range Default Required
1 Weighing
Mode
1
2
3
Single Range
Duel Range
Duel Interval
1 No
2 Trade Mode 0
1
Trade
Industrial
0 No
Use the WMD command to setup the weighing mode of the instrument. This setting
is a fundamental scale build parameter and should be used along with the IAD and
ICR commands before the instrument is calibrated.
Example
Command Response Details
S01;
WMD?; 1,0 CRLF Query current weighing mode
WMD2,1; 0 CRLF Change to dual range, industrial mode
WMD?; 2,1 CRLF Weighing mode is dual range, industrial
TDD1; 0 CRLF Save settings