Rinstrum - 5200 Communications Manual Rev 1.2
Page 16 0025-605-120
2.4.7. CLK: Set Clock
Set the time and date.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
6 Automatic Immediately No
Parameter Details
Parameter Description Range Default Required
1 Hour 0 to 23 - No
2 Minute 0 to 59 - No
3 Second 0 to 59 - No
4 Date 1 to 31 - No
5 Month 1 to 12 - No
6 Year 0 to 99 (or 2000 to 2099) - No
The year may be in 2 digit or 4 digit format. The 5200 will always respond to queries
in 2 digit format.
Example
Command Response Details
S01; Select Instrument 1
CLK?; 5,5,55,22,4,0 CRLF Query date and time
CLK15,40,0,5,2,2003; 0 CRLF Change to 3:40pm 5/2/2003
CLK15,40,0,5,2,3; 0 CRLF Same as above