Rinstrum - 5200 Communications Manual Rev 1.2
0025-605-120 Page 19
2.4.9. CWT: Set Calibration Weight
Set the calibration weight to be used for span calibration. This must be set before
using the LWT; span calibration command.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
1 With TDD1 Immediately No
Parameter Details
Parameter Description Range Default Required
1 Calibration
Weight
2% - 100% of fullscale
weight.
(Send IAD? to read
fullscale setting)
3000 No
Example
Command Response Details
S01; Select Instrument 1
CWT?; 3000 CRLF Query calibration weight setting
CWT4000; 0 CRLF
(Note that weight is sent without
any decimal point. So 400.0 kg
is send as 4000 not 400.0)
Change Calibration Weight to 4000
TDD1; 0 CRLF Save new setting