Rinstrum - 5200 Communications Manual Rev 1.2
0025-605-120 Page 37
2.4.26. PCD: Motion Setting
Enter the full passcode to unlock access to the trade specific settings.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
1 - Immediately No
The FULL passcode is used to control access to trade sensitive parameters. If this
passcode is used it also blocks communications access to these same parameters.
Settings may be read but not written to without first entering the correct passcode
via the PCD command. PCD? Is used to query whether access is currently
blocked. To lock instrument again issue PCD without the passcode.
Parameter Details
Parameter Description Range Default Required
1 Passcode 1 to 999999 - No
Example
Command Response Details
S01; Select Instrument 1
PCD?; 1 CRLF Instrument is locked
PCD1234; 0 CRLF Passcode 1234 accepted
IAD,,,,1; 0 CRLF Put unit in x10 mode.
PCD; 0 CRLF Lock instrument again