Command Reference
R&S
®
ZNB/ZNBT
1020User Manual 1173.9163.02 ─ 54
Suffix:
<Ch>
.
Channel number
<Nr> Segment number
<Nr> Sequence number, defining the order in which the switches/
commands shall be executed.
For every channel, segment and GPIO port, the sequence num-
bers must be consecutive, starting at 1; reusing a sequence
number overwrites a previous voltage setting.
Delays between subsequent settings/command and before the
start of the respective sweep segment can be introduced using
CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:DELay.
<Port> GPIO port number
Parameters:
<Voltage> Range: -6 to +12 V
Increment: 0.005 V
Default unit: V
Manual operation: See "GPIO columns (sweep sequencer table)" on page 525
CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
<Command>
For segmented sweeps this command allows to define the RFFE command(s) to be
executed at the start of each segment.
Command execution on an RFFE interface can be enabled/disabled using
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe]. Complementary GPIO switches
can be defined using CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:GPIO<Port>:
VOLTage.
Suffix:
<Ch>
.
Channel number
<Nr> Segment number
<Nr> Sequence number, defining the order in which the commands/
switches shall be executed.
For every channel, segment and RFFE interface, the sequence
numbers must be consecutive, starting at 1; reusing a sequence
number overwrites a previously defined command.
Delays between subsequent commands and before the start of
the respective sweep segment can be introduced using
CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:DELay
<Bus> RFFE bus interface number
Parameters:
<Command> 3 to 37 hexadecimal digits (0-F), defining the command to be
executed: digit 1 is the slave address, digits 2 and 3 specify the
command number and the remaining digits represent the data
part (up to 17 digit pairs).
SCPI Command Reference