Rohde & Schwarz ZNB40 - Calculate:fmodel:delt<Ph_Pt>:M1 L:measure; Calculate:fmodel:delt<Ph_Pt>:M2 L:measure; Calculate:fmodel:delt<Ph_Pt>:M3 L:measure

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
843User Manual 1173.9163.02 ─ 54
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Length> Length in inches
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Length" on page 619
CALCulate:FMODel:DELT<Ph_pt>:M1L:MEASure
CALCulate:FMODel:DELT<Ph_pt>:M2L:MEASure
CALCulate:FMODel:DELT<Ph_pt>:M3L:MEASure
Measures the full set of S-parameters at the active ports (see CALCulate:FMO
Del:DELT<Ph_pt>:M1L|M2L|M3L[:STATe]).
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Event
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Measure" on page 618
CALCulate:FMODel:DELT<Ph_pt>:M1L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M2L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M3L[:STATe] <Boolean>
Defines the active ports for the 1st/2nd/3rd line Delta-L measurement (CALCu
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
For each measurement, either two or four ports must be active. If the 1-length method
is used (CALCulate:FMODel:DELT<Ph_pt>:METHod M1L), then four ports must be
active for the first line.
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Ports table" on page 618
CALCulate:FMODel:DELT<Ph_pt>:MEASurement <Measurement>
Defines whether single-ended or balanced analysis is used in Delta-L characteriza-
tions.